Samples / NURS FPX 4035 / Assessment 2
Root-Cause Analysis and Safety Improvement Plan
NURS FPX 4035 · Assessment 2
Sample for study and reference. Ace My Assessment is an independent academic support service.
Understanding What Happened | |
What happened? | The nurse working in the cardiac unit has worked there for many years and is accustomed to hearing multiple alarms generated by the bedside monitoring of the patients. The monitoring systems averaged two hundred and ninety-four alarm notifications per patient over multiple 12-hour shifts. Most alarms are either clinically invalid or false alarms. On the night in question, at approximately 9:15 p.m., an alarm sounded for sustained ventricular tachycardia (V-Tach > 100) from the patient's bedside monitor after a heart assessment had been performed on the patient approximately 30 minutes prior to the alarm sounding. The charge nurse, and two of the patient's nurses failed to assess the patient and clear the alarm until after the heart was assessed (i.e. the vitals). Approximately 14 minutes after the alarm, the patient's systolic blood pressure dropped to 81mmHg, and the patient became unconscious until a nursing assistant entered the room and activated the Rapid Response Team upon assessing an emergency situation. As a consequence of this conflict of care, the patient experienced an episode of hypoxia requiring mechanical ventilation for a temporary period of time, remained in ICU for an additional 9 days and experienced permanent mild cognitive impairment as a direct result of prolonged cerebral hypoperfusion. The spouse of this patient submitted a complaint in writing to the committee regarding the patient's avoidable harm due to the desensitization to alarms and the lack of a structured protocol to respond to alarms. The resulting investigation verified that the patient had a critical alarm that had been on the central monitor and on all three bedside monitors for at least 14 minutes and had to have been clinically assessed by nursing staff prior to the alarm being cleared. |
Why did it happen? | Human Factors Due to a long-term exposure to continuous non-actionable alert notifications, nursing staff across both day and evening shifts developed a habitual pattern of dismissing monitor alarm signals. The nurse primarily responsible for Patient B's bedside care acknowledged that she had turned off the alarms on Patient B's telemetry for over 30 of her current shift of duty and that the critical alert notification for atrial fibrillation (ventricular tachycardia) sounded identical in tone versus previous false alerts (Yang et al., 2024). The charge nurse who supervised six patients who were being monitored simultaneously during Patient B's incident did not perform central telemetry monitoring; therefore, there was no centralized oversight of alarms that had gone off during their busiest time during peak evening hours. Cognitive load theory also provides insights into this incident. As hospital staff are exposed to higher than normal levels of stimuli (to include false alarms) that do not represent a clinical need for their response, they will continue to divert their attentional resources away from providing oversight of alarm signals, and there will be a deterioration in their ability to recognize true emergency signals as the amount of time they are exposed to meaningless stimuli increases. System Factors The cardiac ICU at Northridge Memorial did not have specific Protocols for Alarm Management specific to each Unit, nor did it have calibration settings that were specifically individualized to each patient based on his or her Physiological Baseline. All monitoring equipment was set up at the Manufacturer Default Settings, which were neither reviewed nor adjusted upon Admission or Daily Rounds; Studies show that this results in substantially higher rates of non-actionable alarms than Individualized Threshold Programming (Güner & Üçüncüoğlu, 2026). The Monitoring Infrastructure of the Hospital did not include Intelligent Filtering Algorithms to distinguish between Artifact and True Arrhythmias, nor did the Monitoring Infrastructure have Tiered Escalation Technology to automatically notify Secondary Responders if the Primary Nurse failed to Respond to an Alarm within a Defined Response Timeframe. There was no formal Alarm Management Committee or Interdisciplinary Oversight Committee in place to regularly review Alarm Event Data, identify High-Burden Units, or implement Corrective Threshold Adjustments. Organizational Culture The cardiac ICU's institutional norms reinforced the idea that silencing alarms was a routine nursing practice and that it was an efficient way for nurses to provide care–not a nursing practice that could jeopardize the safety of patients. In post-incident nurse surveys, nurses felt tremendous informal pressure to silence alarms quickly (to limit disturbances to other patients) and to keep the unit noise level within an acceptable range; the thoroughness of alarm response was secondary to the efficiency of workflow (Fouad et al., 2025). The senior nursing leadership had received multiple prior reports indicating that alarm fatigue was becoming a significant safety issue, but had deferred action because of budgetary constraints and competing quality improvement objectives. The lack of evidence that senior nursing leadership was committed to reforming alarm management practices sent an implicit message to the organization that alarm burden was an anticipated condition associated with nurses' occupation, not a systemic hazard that could be corrected. Society/Culture The cultural environment of healthcare workers all over the world, impacted by professional norms, has created alarm exhaustion in nurses by discouraging reporting of alarm-related near-misses or raising concerns about monitoring practices with the supervising staff. Nurses trained in high-alarm-volume environments have accepted the practice of rapid silencing alarms as evidence of efficient clinical practice, not as a potential safety risk, and institutional on-boarding procedures have not provided alarm prioritization or response alternative/expanded frameworks (Arkilic et al., 2024). Additionally, the hierarchical dynamics of the ICU made it difficult for junior nursing staff to question alarm response practices or advocate for protocol revisions when they perceived their own competency to be implicitly jeopardized. Additionally, due to the widespread nature of alarm fatigue at many hospitals across the USA, there is a reduced urgency to address alarm fatigue in any given hospital. |
Was there a deviation from protocols or standards? | The hospital monitoring policy outlined that nurses must perform a direct check of all patients within 90 seconds of any sustained arrhythmia alarm being triggered. However, documentation shows no response to the Ventricular Tachycardia (VT) alarm activated on Patient B's monitor until 14 minutes after the alarm had first sounded. The Joint Commission’s National Patient Safety Goal (NPSG) 06.01.01 requires health care organizations to develop and implement a written policy detailing how to manage their clinical alarm systems, which includes protocols for responding to alarms, customizing the threshold parameters for alarms, and regulating the escalation process in the event there is no response by the nurse within the specified timeframe. The hospital did not have these operational policies/protocols in place on this unit (Nyarko et al., 2022). The hospital’s equipment management policy stated that threshold parameters for the monitor should be assessed and configured to the individual patient's needs within 24 hours of him/her being admitted; however, upon audit of Patient B’s record, he was placed on factory-default monitor settings for the entire 3 days he was in the ICU without documentation of him being assessed and configured to the monitor. The hospital's policy also required that all critical alarm events be documented in the electronic medical record (EMR) with the nurse’s assessment and rationale for changing the alarm; however, a review of the EMR for the past 14 alarm events on Patient B confirms that the nurse did not meet this requirement. In addition, there were 43 instances where bedside nurses silenced the alarm between 1900-2300, but the alarm escalation protocol requiring charge nurses to conduct an independent assessment of patients after bedside nurses have silenced a predetermined number of alarms within a specified time frame was not triggered or followed for Patient B’s record (Xu et al., 2025). |
Who was involved?: | Staff During Patient B's evening shift, the nurse assigned to care for him was responsible for responding to alarms and checking on his needs while also caring for five other patients with alarms and family communication, administering medications, and assessing post-procedural wounds. During the incident period, the charge nurse had oversight of the unit and some responsibility for monitoring the central station; there was no individual telemetry observer assigned for the shift (Dee et al., 2022). Two other bedside nurses working with patients adjacent to Patient B's bed were aware of the repeated alarm patterns from Patient B's unit. They had both cleared audible alarms without assessing Patient B. This behavior of unintentional dismissal of an alarm occurred regularly throughout the unit as a result of the collective decision by the nursing staff. The nursing assistant responsible for identifying Patient B’s deterioration and activating the Rapid Response Team was not credentialed to review telemetry data or answer alarm communications; however, acted appropriately when the need for assistance was visually obvious. Supervisors and Managers Although the ICU nursing manager had access to unit-level alarm event reports generated on a monthly basis by the biomedical engineering department, she had failed to take any action with the data indicating that the cardiac intensive care unit (ICU) has an average of more than 250 alarm events per patient per day (a volume well above published benchmarks in peer-reviewed research indicating that this volume correlates with both alarm fatigue and adverse events) (Alsuyayfi & Alanazi, 2022). The nursing director has been informed of a previous patient safety committee's recommendation to establish an interdisciplinary alarm management task force, yet there have been no staff time allocations or budgetary resources provided for the task force to be established prior to this incident. The patient safety officer had documented three near misses that occurred on the same unit that referenced delayed response to alarms over an eight-month period, but he had labeled each of those issues as an individual personnel performance problem rather than identifying a systemic problem that requires an organization-wide intervention. |
Was there a breakdown in communication? | Interdisciplinary Communication During the afternoon telemonitoring of Patient B, the cardiac monitor tech had indicated in the internal monitoring log that there were unusual alarm densities/variances within the telemetry system. However, this issue was never communicated to the bedside nurse staff or included in the clinical handoff during shift change (Chromik et al., 2022). The cardiologist who reviewed Patient B's post-procedural status during the early afternoon also recorded added arrhythmia risk factors in the physician’s progress note area of the electronic medical record. However, there was no systematic means for notifying nursing staff of the newly updated monitoring parameters requiring increased monitoring due to the implementation of a new drug initiated by pharmacy in the early afternoon, which also included a documented proarrhythmia potential, but was only communicated via the medication administration record and did not generate a nursing alert or initiate a change to alarm threshold recommendations. Patient-Provider Communication During afternoon care, Patient B expressed to the nursing assistant that his monitor alarm was going off frequently and that the sound was preventing him from being able to rest. He asked if the alarms indicated something wrong with his condition. This concern was not reported by the nursing assistant to the RN nor recorded in any of Patient B's medical records as it was a patient-reported finding that could have warranted earlier clinical review (Piri et al., 2022). During admission, the admitting RN did not provide sufficient education to Patient B and his family regarding what the monitoring alarms meant; when they should activate their own call light if the alarms are sounding; or what the expected patient and family response is to alarms that sound and remain unanswered. Education regarding alarm significance would have provided Patient B's family with an added level of safety redundancy that may have prompted his family to activate nursing staff earlier in the critical arrhythmia phase. |
What were the contributing factors? | Physical Environment The cardiac ICU was an open bay unit with 12 monitored patient stations; all alarm sounds were heard throughout the entire area at the same time, generating continuous background noise levels described by staff interviews to be psychologically draining and causing cognitive difficulties (Salameh et al., 2024). There was no acoustic zoning, sound-damping, or localized notification technology for alarms from the units. The central monitoring location was located at one end of the unit while nurses were providing care to patients in bays that were much further away. This made it very difficult for the nurse to check the details of an alarm event at the central monitor and made it less practical for the nurse to verify multiple eruptions of alarm during an emergency using central monitoring. The emergency call system and overhead paging alone averaged fourteen (14) different types of interruptions per hour that prevent nursing staff from being able to pay close attention to monitoring their alarm stream. Staffing Levels During the evening hours on the cardiac intensive care unit (ICU), the nurse-to-patient ratio of 1:6 was greater than the recommended 1:2 or 1:3 standard for monitored intensive care. This resulted in every nurse having to coordinate multiple responsibilities (e.g., alarms, medication, documentation, and family communication) for twice the number of patients, which is considered unsafe for high-acuity cardiac monitoring (Carelli, et al., 2022). Due to cost-cutting measures instituted six months prior to the event, no telemetry monitoring technicians were assigned to the evening shift, eliminating a secondary layer of surveillance that had previously provided central monitoring oversight without increasing the nursing workload at the bedside. Additionally, budgetary constraints decreased the overlap period from thirty minutes to eight minutes between day and evening shifts, thereby compressing the amount of time available for an adequate clinical handoff and review of the monitoring status of high-acuity patients. Training and Competency The review of training records indicates there was no unit-specific alarm management competency development curriculum in place for nursing staff in the cardiac ICU, and training related to alarm response included only a 15-minute module in the general orientation program for the hospital that taught nurses how to operate the equipment but did not teach alarm prioritization strategies (such as recognition of fatigue and escalation protocols) (Movahedi et al., 2023). Additionally, nursing staff received no training on customizing the alarm thresholds or using clinical criteria to determine appropriate ranges of escalation for individual patients with complex histories of cardiac disease. Lastly, the competency assessment for nurses working in the cardiac ICU to evaluate competency regarding alarm management does not include direct observation of how each nurse responds to alarms or how effectively they use established frameworks to prioritize concurrent alarm events; therefore, no deficiencies associated with alarm management practices have been identified via existing methods of performance evaluation. |
Did organizational policies or procedures play a role? | Policy Compliance The clinical alarm management policy established by the hospital (and approved to meet the Joint Commission standards) has designated procedures regarding individualizing alarm thresholds, defining response times, establishing escalation protocols, and requiring documentation of procedures; however, nurses within the cardiac ICU reported that these processes were not being implemented consistently (Albanowski et al., 2023). Following an incident within the cardiac ICU, nursing staff who were interviewed reported that the majority had never been made aware of the response time requirement, nor had they ever had to review and signify review of the alarm management policy as part of their annual competency review. Furthermore, the requirement for alarm threshold parameters being documented as reviewed within 24 hours of admission to a patient was not configured as a required data field within the electronic health record; therefore, non-compliance was not visible to supervisors for the purpose of monitoring compliance with the policy. Policy Clarity The hospital’s institutional alarm management policy consists of thirty-two pages that comprise a number of different areas of policies in the hospital's policy library. Additionally, each unit would have specific requirements listed in the general alarm section of the policy, which would be very difficult for someone without thorough familiarity with the document structure to find (Lewandowska et al.; 2023). An important note, however, is that the policy fails to present different types of alarm responses for those patients who have documented arrhythmia risk factors, has no limit to the number of times a patient can silence their monitor alarm in a single shift before the Charge Nurse/Manager is notified and does not specify the level of accountability for the Central Monitoring Specialist during busy nursing periods. There were no concise, unit specific alarm management job aids available to the nurse, no reference cards or electronic health record (EHR) decision support options available for the nurse, and therefore the nurse must rely on their memory in a highly distractible clinical environment to interpret and implement policy. |
Was there a failure in monitoring or surveillance? | Post-Discharge Follow-up A quality monitoring program in the hospital was not currently maintaining an organized way of tracking how many times alarms are sounded, how often they are silenced, and whether or not they were silenced in compliance with hospital standards on a unit-by-unit basis; therefore, the cardiac ICU’s significant number of alarms sounded daily was not recognized as an indicator of quality prior to the incident with Patient B (Huo et al., 2022). The data from alarms generated automatically through monitoring devices is collected and maintained by the biomedical engineering department, but the department does not routinely review this data to determine the clinical safety implications or share the data with the nursing administration, hospital patient safety officers, or hospital quality improvement coordinators in such a way that would create enthusiasm for corrective actions. In relation to the incident reporting system, there is no organized method for recording alarm fatigue-related near-miss events, so alarm fatigue-related safety concerns are generally recorded as “communication” or “monitoring,” which limits the ability to conduct focused evaluations and develop corrective action plans. Quality Monitoring The cardiac ICU did not establish baseline metrics for alarm management performance, which hinders both identifying alarm response compliance that is trending downward and comparing the unit's performance to that of other units nationally, for example to American Association of Critical-Care Nurses (Sendelbach & Funk, 2023). The quality improvement department did not monitor alarm management as a patient safety indicator, despite the fact that The Joint Commission has defined alarm safety as a National Patient Safety Goal since 2014. As such, there appears to be considerable disconnect between regulatory back-and-operational performance measurement at this institution. Staff satisfaction surveys consistently indicated that alarm volume and alarm fatigue are major sources of occupational stress over three consecutive annual administrations; however, there has not been any effort to connect this data to an analysis of patient safety risks or use this data to initiate alarm management quality improvements (Nyarko et al., 2023). |
What can be learned to prevent recurrence? | Lessons Learned This adverse event illustrates that the systemic danger of alarm fatigue, in high-acuity monitoring environments, is an institutional hazard, not a failing on the part of individual performers, and that developing sustainable safety improvements requires an acknowledgment of the technological, environmental, staffing and cultural conditions that facilitate and allow alarm desensitization to develop and flourish (Li et al., 2024). It also demonstrates the necessity of operationalizing the Joint Commission's alarm safety guidelines through the establishment of unit-specific protocols with clear accountability, measurable compliance standards and regular supervisory reviews, rather than merely documenting for the purpose of demonstrating compliance with regulations. The case also underscores that patient and family participation in alarm safety, when combined with education to nurses and patients regarding the significance of monitoring and appropriate responses to alarms, provides an underutilized redundancy layer that may contribute significantly to risk reduction. Quality Improvement The implementation of a comprehensive package for alarm management will address vulnerabilities in alarm management based on lessons learned from this event. Creating an easily accessible alarm management committee comprised of nursing, medical, biomedical, and quality improvement team members will create an ongoing governance structure that is able to review alarm event data regularly and continuously address alarm burden patterns and ensure adherence to established protocols. This integration of alarm management performance metrics into the current quality monitoring dashboard as well as linking those metrics to nursing leadership performance expectations will assist in creating accountability for alarm safety across both the unit and institutional organizations. |
How can patient safety be enhanced? | Risk Mitigation Making individualized threshold alarm reviews mandatory when documenting and setting up the admitting electronic health record would help reduce the number of non-actionable alarms (Sinno et al. 2022) with the use of patient-specific physiologic profiles to set monitoring parameters, rather than using established generic or default values. Implementing an escalatory notification system that automatically alerts the charge nurse pagers or provides a unit-wide visual alert after an established response time period for critical alarms would eliminate the possibility of having a single point-of-failure that prevented Patient B's irregular heartbeat from being addressed by the nurse for 14 minutes. Implementing a dedicated telemetry role on each shift, staffed by a nurse or technician dedicated solely to monitoring central telemetry, could provide independence from the individual nurse workload when monitoring patients through telemetry. Education and Training Creating a thorough alarm management training program that is tailored for cardiac ICUs will provide a solid reading and understanding of what constitutes a proper monitoring procedure (Nyarko et al., 2023). This can be done by providing sufficient classroom training related to customizing alarm thresholds, conducting simulation exercises designed to practice the prioritization of alarms, and using standardized assessments to evaluate nurses' responses to alarms. Annual retraining through direct observation of nurse responses to alarms in combination with using historical alarm data from the unit will help ensure that our standards are enforced and ensure that any nurses in need of additional training can be identified before an untoward event occurs. Providing information about alarm systems to the patient and family members who are admitted to monitored units will provide a patient safety mechanism that is helpful, yet presently not used frequently. The information to be provided includes information on why alarms exist; instructions for independently activating the nurse call system; instructions for appropriately responding to alarms that are delayed in response; and the significance of alarms. Reporting and Feedback By means of an incident reporting system to collect data on alarm response failures before they cause any harm to the patient, it is possible to create a dedicated alarm fatigue near miss reporting category. This could allow for a proactive approach to improve safety (Ruppel et al., 2023). To continue to have an ongoing organizational focus on alarm safety, it would be beneficial for all nursing staff, charge nurses, and unit managers to receive a monthly display of unit level alarm performance dashboards that provide summary information pertaining only to the volume of alarms generated, number of times an alarm has been silenced or responded to in compliance with the author's recommendation, and the threshold individualized rate of compliance for each alarm. A psychologically safe reporting environment would encourage nurses to report concerns about the alarm systems, flag patients who are at a high burden of alarm noise, and request support with managing these types of alarms without fear of blame attribution provide the foundation for the development of a culture that will support continuous improvement towards the furtherance of alarm safety. |
Root Cause(s) to the issue or sentinel event?
Root Cause – the most basic reason that the situation occurred | Contributing Factors – additional reason(s) that clearly made a situation turn out less than ideal | HF-C | HF-T | HF-F/S | E | R | B |
Absence of individualized alarm threshold management protocols and dedicated telemetry surveillance roles in the cardiac ICU | 1Factory-default alarm parameters applied uniformly; no structured process for patient-specific threshold review; no secondary escalation technology | X | X | ||||
Normalization of alarm silencing behavior and absence of accountability structures for alarm response compliance | 2No unit-level alarm management competency training; no documentation requirement for alarm response rationale; institutional culture permitting routine alarm dismissal | X | X | X | |||
Inadequate nurse staffing ratios and elimination of dedicated telemetry monitoring technician roles | 3One-to-six nurse-to-patient ratios exceeding ICU safety standards; no dedicated surveillance technician; compressed shift overlap eliminating comprehensive handoff | X |
HF-C = Human Factor-communication HF-T = Human Factor-training HF-F/S = Human Factor-fatigue/scheduling
E= environment/equipment R= rules/policies/procedures B=barriers
Application of Evidence-Based Strategies
Identify evidence-based best practice strategies to address the safety issue or sentinel event.
Numerous studies have demonstrated that the use of alarm management programs that are structured can help reduce patient harm associated with alarm fatigue in high-acuity clinical environments. To support this conclusion, Nyarko et al. (2023) conducted a systematic review of comprehensive alarm management programs, which included: daily electrode maintenance; individualized threshold programming; and structured RN educational interventions. They found that alarm management programs using these comprehensive approaches were able to reduce total alarms by thirty to forty percent and increase the speed of RN response to significant clinical alarms across multiple acute care settings. In addition, Salameh et al. (2024) concluded that failure to individualize monitoring parameters was the most readily modifiable cause of alarm fatigue, and that unit-based protocols to routinely review thresholds reduced non-actionable alarms by thirty-seven percent within ninety days of implementation. In the cardiac monitoring environment, Xu et al. (2025) demonstrated that the implementation of secondary escalation systems that route unacknowledged critical alarms to charge nurses resulted in a decrease of average alarm response times from eleven minutes to less than ninety seconds in a prospective trial study. This finding is of particular significance to Patient B's adverse event, whereby a visible critical alarm did not receive a response for a period of fourteen minutes. Ruppel et al. (2023) found that the use of structured alarm management competency education utilizing simulation increased compliance with alarm response policies from fifty-one percent to eighty-nine percent at the six-month follow-up, demonstrating a significant evidence of behavior change from an educational intervention in the area of alarm management. Xu et al. (2025) identified that through patient and family education on the importance of alarms and the alarm escalation protocol, RN notification of a patient's deterioration decreased by twenty-two percent and noted that these educational interventions were an underutilized safety strategy. Taken collectively and in their entirety, this body of literature supports a multi-faceted approach to alarm management that simultaneously incorporates the technological, educational, staffing structure, and patient engagement components of alarm management. |
Explain how the strategies could be applied in the safety issues or sentinel events you have identified.
The cardiac ICU will utilize the alarm bundling protocol that was developed by Arkilic et al. (2024) through a daily alarm review and recording procedure that each bedside nurse completes for their assigned patients at the beginning of each nursing shift, including verification of alarm threshold reviews as a mandatory component of electronic medical record documentation, thus directly addressing the primary root cause of Patient B's adverse event by requiring nursing staff to evaluate Patient B's arrhythmia risk status and customize his monitoring parameters to fit his particular clinical scenario as opposed to utilizing factory default settings. Incorporation of the secondary escalation technologies proposed by Dee et al. (2022) will require notification to charge nurses via pager through integration of existing monitoring systems, requiring configuration modifications in the biomedical engineering department and negating the need for capital equipment purchase, and also create an automated safety net for all critical alarms that are not acknowledged at the bedside. The simulation-based alarm training curricula proposed by Huo et al. (2022) will be tailored to the cardiac ICU context through development of unit-specific patient scenarios related to recognition of arrhythmia, alarm prioritization during concurrent alert conditions, and structured decision making related to escalation, with assessment of competency through direct observation of alarm response behavior during unannounced competency evaluation. Alarm event data reviewed by Nyarko et al. (2022) will guide the establishment of unit-level performance dashboards that will track alarm volume per patient daily, alarm silence rates, compliance with individualized alarm thresholds, and response time distributions, providing nursing leadership with actionable real-time information for identifying developing alarm burden trends prior to producing negative patient outcomes. Patient and family education recommendations outlined by Sinno et al. (2022) will be part of the standard ICU admission nursing assessment, through provision of structured education sessions regarding the importance of monitoring alarms, circumstances in which the patient or their family members may need to independently activate the call system, and expectations for nursing staff's response to alarms. |
Safety Improvement Plan
One for each Root Cause/Contributing Factor from above | E / C / A Choose one | |
1 | Cardiac ICU must implement a broadly applied alarm management bundle. This includes an individualized alarm threshold review and documentation of every patient at the start of every shift, as well as integrating technology that sends all critical alarms that have not been acknowledged to charge nurse devices after 90 seconds. There must also be a dedicated telemetry monitoring role on each shift staffed exclusively by a registered nurse or trained technician to perform surveillance from central telemetry monitors. | E |
2 | Create and implement a required competency program for cardiac ICU staff for managing alarming devices in accordance with established criteria by incorporating educational content on establishing individualized threshold criteria for alarms; training the use of alarms through simulation in concurrent alarm conditions; structured and evaluated experience in making clinical decisions to escalate care for patients with an alarm; and assessment of individual staff competencies with feedback after all nursing staff on the unit have completed the competency program within the first 90 days and then at least once per year after that. | C |
3 | Develop and implement a program for tracking and reporting alarms at the unit level, including an annual report for alarm event volume, alarm silencing rate, setting of individualized high/low alarm thresholds, critical alarm response times, and monthly reporting of this data to all nursing staff, charge nurse and unit management. Additionally, create a structured, interdisciplinary alarm management committee that meets quarterly to evaluate trends and make changes to protocols or policies as necessary. | E |
E = eliminate (i.e. piece of equip is removed, fixed or replaced.)
C = control (i.e. additional step/warning is added or staff is educated/re-educated)
A = accept (i.e. formal or informal discussions of “don’t let it happen again” or “pay better attention” but nothing else will change and the risk is accepted)
Describe any new processes or policies and/or professional development that will be undertaken to address the root cause(s).
A cardiac Intensive Care Unit (ICU) Alarm Management Policy is being created in order to replace the existing hospital-wide alarm management policy. The cardiac ICU Alarm Management Policy will be unit-based and will reflect the operational characteristics and patient population complexities of a high acuity cardiac monitoring environment. Specific mandatory procedures for individualizing the alarm thresholds will be included in this new alarm management policy. These procedures will include procedures for establishing alarm parameter ranges for patients at risk for experiencing an arrhythmia; requirements for response time for each level of alarm classification; and a maximum number of times per patient, per shift, that the charge nurse must be notified if the alarm continues to be silenced by the same patient. In addition, the central monitoring station will have specific accountability for monitoring an assigned number of patients for each shift (Nyarko et al., 2023). Electronic health records will be modified to require that documentation concerning the individualization of alarm thresholds will be available in the nursing shift assessment as a part of the nursing workflow. Decision support prompts will provide guidance to nursing staff in the process of using the individualization criteria for alarm thresholds by using criteria pre-populated from the patient's medical record. A comprehensive professional development program will be created in collaboration between the Education Department and the ICU. The program will consist of a four-hour initial training with simulation activities to practice prioritizing alarms and developing a competency checklist to assess the nurse's response to alarms during direct observation of alarms in the practice setting, as well as a one-hour annual revalidation program, which will cover a review of the unit-level alarm performance data and updates on the protocols. The nurse manager will be responsible for reviewing the monthly alarm performance dashboard data and creating an action plan if performance metrics fall below the benchmarks established by the ICU. In addition, an interdisciplinary alarm management committee, including nursing, cardiology, biomedical engineering, pharmacy, and quality improvement representatives, will meet quarterly to review the alarm management performance of the alarm management policy and to provide evidence-informed updates to the protocol (Salameh et al., 2024). |
Provide a description of the goals or desired outcomes of the actions listed above, along with a rough timeline of development and implementation for the plan.
SMART Goals Within six months after implementing the alarm management bundle, there will be a reduction in the total number of non-actionable alarms from 294 alarms per patient per day down to less than 150 alarms per patient per day in the cardiac ICU. To measure this success, the data will be obtained from the biomedical engineering analysis of alarm history. Baseline data for this metric was collected during the first month of the project. There will be an increase in compliance to the individualized alarm thresholds documented in the electronic health record (EHR) during shift assessments of ED patients in the cardiac ICU. Target compliance rates are: 95% by month 4 of project implementation and 98% by month 7 of project implementation. Compliance data will be tracked weekly through EHR audit reports produced by the Quality Improvement department. There will be a reduction in the mean response time to critical alarms from the current baseline of 14 minutes down to less than 90 seconds on 90% of critical alarm events within 5 months of project implementation. Actual response times for critical alarms will be documented in the monitoring system; all alarm management committee members will review the monitoring data monthly. All cardiac ICU nursing staff will successfully complete alarm management training and be competency assessed at 100% within 90 days after the curriculum is finalized. Evidence of successful completion of competency assessment will be determined via direct observation of nurse performance on alarm prioritization and threshold individualization skills and will result in 90% of staff passing during the initial evaluation. Documented evidence of this will be available in the nursing education department learning management system. No preventable adverse patient event attributable to delayed response to critical alarms will occur in the cardiac ICU for the first 12 months following full implementation of the alarm management bundle. Data will be reviewed through monthly analysis of incident reports/causal analysis as it relates to patient safety — specifically focusing on alarm management contributing to incidents. Implementation Timeline Phase 1: Assemble a cross-disciplinary team for alarm management with participation from nurses in the cardiac ICU who work different shifts, charge nurses, unit manager, cardiologists, biomedical engineering, clinical informatics, pharmacy, and the patient safety department. Conduct a comprehensive scan of baseline measurements: alarm event volume analysis, staff knowledge survey, policy gaps assessment, and EHR workflow analysis. Identify 2-3 nurse champions on the unit to assist with peer support for implementation. Phase 2: Finalize the alarm management policy with unit-specific threshold individualization criteria, response time requirements, escalation protocols, and accountability assignments; complete design of the EHR workflow for mandatory threshold documentation fields and clinical decision support prompts; develop a simulation-based training curriculum that includes patient scenarios, competency checklists, and assessment rubrics; procure and configure technology that enables notification of secondary escalation for integration with current monitoring technology. Phase 3: Test the revised alarm bundle on 2 cardiac ICU bays with a voluntary nursing staff; start competency training for the pilot participants; turn on the EHR threshold documentation requirements for the pilot beds; evaluate the secondary escalation notification technology's performance and address any technical issues; obtain weekly feedback from the pilot nursing staff and revise the protocols based on findings from daily operations. June through August: Complete the rollout of the alarm management bundle to all ICU beds; conduct competency training for the nursing staff with evaluator observations; conduct a weekly alarm performance review in collaboration with the charge nurses; create a new dashboard for nursing leadership to review each unit's monthly alarm performance metrics and reports; include patient/family education on alarms within the routine nursing assessment of the patient upon admission to the ICU; evaluate the effectiveness of the alarms by comparing current alarm volume and response times to baseline measures. September through November: Conduct the evaluation of six months of alarms events to determine the event rates, critical alarm response time, compliance with individualized alarm thresholds, and near-miss report frequency against both baseline measures and the SMART goals. Present the outcomes of the study back to both hospital patient safety and nursing leadership; refine the alarm management protocols based upon study findings as well as ongoing staff input and feedback; provide remedial assistance to staff that were not competent based on their evaluation of their performance; provide recognition to high performance shifts and nurses through formal recognition programs. December through February: Completed a twelve-month comprehensive evaluation of the alarm management program with a detailed analysis of alarm volume reduction, adverse event rates, staff competency, and estimated cost-savings due to complications that were avoided as a result of effective alarm management; present the overall program results and recommendations for sustaining the program to the chief executive officer and board's quality committee; create an implementation guideline for replicating the alarm management program on step-down and general medical/surgical patient care area; develop an annual competency verification process and ongoing governance for alarm management within the organization; identify quality improvement results for either external publication or conference presentation. |
Existing Organizational Resources
The cardiology monitoring system in place at the hospital is current and does not have any capacity to escalate alarms (secondary escalation capability) or provide data analytics filtration processes (intelligent filtering processes), however, the software system is already integrated into the hospital’s technology environment and is capable of integrating with the alarm notification routing and threshold management workflows as a result of compatibility verified by biomed engineering. This represents the greatest short-term opportunity to reduce risk of harm by providing additional means for alarm threshold individualization as well as providing additional means to activate or de-activate alarm escalation notification based upon current compliance to established workflows, without having to purchase additional capital equipment for alarm management (Movahedi et al. 2023) and; this also provides an opportunity to use this existing alarm management technology that can further avoid procurement delays and additional resources required to train staff appropriately when purchasing new alarm management equipment, while also supporting the primary evidence-based strategies for risk mitigation and preventing incidence of alarm fatigue that includes secondary surveillance means and alarm individualization capabilities. The hospital's education and simulation center for nursing provides an actual practice environment that is fully equipped to provide high-fidelity simulation of the patient and system/environmental conditions required for the training program being proposed, including the alarm saturation and concurrent prioritization conditions integral to competency training. The nursing education coordinator of the cardiac ICU has developed a working relationship with the simulation center faculty such that a collaborative process can develop the training curriculum utilizing both the clinical expertise of the cardiac ICU nursing education coordinator and the educational methodologies of the simulation center faculty at the same time. In addition, since the simulation center is established as the standard for competency assessment, the use of performance measurement tools that have been developed will enable the development of alarm management competency evaluation tools for the training program without the additional time or cost of developing evaluation tools from the beginning (Carelli et al., 2022). Therefore, the development process for the educational component of the training program will be considerably shortened and the costs associated with the development of the educational component of the training program will be minimized while ensuring that the training is delivered in a consistent, assessed manner and the results of the training are able to provide documentation of being competent for regulatory requirements and accreditation processes. |
References
Albanowski, K., Burdick, K. J., Bonafide, C. P., Kleinpell, R., & Schlesinger, J. J. (2023). Ten years later, alarm fatigue is still a safety concern. AACN Advanced Critical Care, 34(3), 189–197. https://doi.org/10.4037/aacnacc2023662
Alsuyayfi, S., & Alanazi, A. (2022). Impact of clinical alarms on patient safety from nurses’ perspective. Informatics in Medicine Unlocked, 32, e101047. https://doi.org/10.1016/j.imu.2022.101047
Arkilic, L. Z., Hundt, E., & Quatrara, B. (2024). Critical care alarm fatigue and monitor customization: Alarm frequencies and context factors. Critical Care Nurse, 44(2), 21–30. https://doi.org/10.4037/ccn2024797
Carelli, L., Terzoni, S., Destrebecq, A., Formenti, P., Soumahoro, F., Esposito, A., & Ferrara, P. (2022). Alarm fatigue in nurses working in intensive care units: A multicenter study. Work, 72(2), 01–06. https://doi.org/10.3233/wor-210552
Chromik, J., Klopfenstein, S. A. I., Pfitzner, B., Sinno, Z.-C., Arnrich, B., Balzer, F., & Poncette, A.-S. (2022). Computational approaches to alleviate alarm fatigue in intensive care medicine: A systematic literature review. Frontiers in Digital Health, 4. https://doi.org/10.3389/fdgth.2022.843747
Dee, S. A., Tucciarone, J., Plotkin, G., & Mallilo, C. (2022). Determining the impact of an alarm management program on alarm fatigue among ICU and telemetry RNs: An evidence based research project. SAGE Open Nursing, 8, e10987. https://doi.org/10.1177/23779608221098713
Fouad, M. R., Al-Metyazidy, H. A., El-Masry, M. M., Gomaa, N. S., & Abd Elhafez, N. G. E. (2025). Implementing safety bundle regarding alarm fatigue on nurses’ perception, practices and patients’ clinical outcomes in cardiac care units. Sohag Journal of Nursing Science, 4(7), 95–109. https://doi.org/10.21608/sjns.2025.396259.1077
Güner, Y., & Üçüncüoğlu, M. (2026). Identifying the alarm fatigue status of nurses working in cardiac intensive care units: A mixed-design study. American Journal of Critical Care, 35(1), 32–43. https://doi.org/10.4037/ajcc2026265
Huo, J., Wung, S.-F., Roveda, J., & Li, A. (2022). Reducing false alarms in intensive care units: A scoping review. Exploratory Research and Hypothesis in Medicine, 0(000). https://doi.org/10.14218/ERHM.2022.00026
Lewandowska, K., Mędrzycka-Dąbrowska, W., Tomaszek, L., & Wujtewicz, M. (2023). Determining factors of alarm fatigue among nurses in intensive care units—A Polish pilot study. Journal of Clinical Medicine, 12(9), 3120–3120. https://doi.org/10.3390/jcm12093120
Li, B., Yue, L., Nie, H., Cao, Z., Chai, X., Peng, B., Zhang, T., & Huang, W. (2024). The effect of intelligent management interventions in intensive care units to reduce false alarms: An integrative review. International Journal of Nursing Sciences, 11(1), 133–142. https://doi.org/10.1016/j.ijnss.2023.12.008
Movahedi, A., Sadooghiasl, A., Ahmadi, F., & Vaismoradi, M. (2023). Smart care for dealing with nurses’ alarm fatigue in the intensive care unit. Journal of Nursing Scholarship, 55(4). https://doi.org/10.1111/jnu.12870
Nyarko, B. A., Nie, H., Yin, Z., Chai, X., & Yue, L. (2022). The effect of educational interventions in managing nurses’ alarm fatigue: An integrative review. Journal of Clinical Nursing, 32(13-14). https://doi.org/10.1111/jocn.16479
Nyarko, B. A., Yin, Z., Chai, X., & Yue, L. (2023). Nurses’ alarm fatigue, influencing factors, and its relationship with burnout in the critical care units: A cross-sectional study. Australian Critical Care, 37(2). https://doi.org/10.1016/j.aucc.2023.06.010
Piri, H., Huh, W. T., Shechter, S. M., & Hudson, D. (2022). Individualized dynamic patient monitoring under alarm fatigue. Operations Research. https://doi.org/10.1287/opre.2022.2300
Ruppel, H., Dougherty, M., Bonafide, C. P., & Lasater, K. B. (2023). Alarm burden and the nursing care environment: A 213-hospital cross-sectional study. BMJ Open Quality, 12(4), e002342. https://doi.org/10.1136/bmjoq-2023-002342
Salameh, B., Abdallah, J., Alkubati, S. A., & ALBashtawy, M. (2024). Alarm fatigue and perceived stress among critical care nurses in the intensive care units: Palestinian perspectives. BMC Nursing, 23(1). https://doi.org/10.1186/s12912-024-01897-x
Sinno, Z.-C., Shay, D., Kruppa, J., Klopfenstein, S. A. I., Giesa, N., Flint, A. R., Herren, P., Scheibe, F., Spies, C., Hinrichs, C., Winter, A., Balzer, F., & Poncette, A.-S. (2022). The influence of patient characteristics on the alarm rate in intensive care units: A retrospective cohort study. Scientific Reports, 12(1). https://doi.org/10.1038/s41598-022-26261-4
Xu, D., Liu, F., Ding, X., Ma, J., Suo, Y., Peng, Y.-Y., Li, J., & Fu, X. (2025). Exploring ICU nurses’ response to alarm management and strategies for alleviating alarm fatigue: A meta-synthesis and systematic review. BMC Nursing, 24(1). https://doi.org/10.1186/s12912-025-03084-y
Yang, J. K., Su, F., Graber-Naidich, A., Hedlin, H., Madsen, N., DeSousa, C., Feehan, S., Graves, A., Palmquist, A., Cable, R., & Kipps, A. K. (2024). Mitigating alarm fatigue and improving the bedside experience by reducing non-actionable alarms. The Journal of Pediatrics, 114278–114278. https://doi.org/10.1016/j.jpeds.2024.114278